RAS-08REV A.2RAS

Sterilization Workflow & Sterile Field

Mixed reprocessing strategy: autoclave for mechanical hardware, STERRAD VHP for sensor head, gamma for single-use cartridges. Validated SAL 10⁻⁶.

Engineering DescriptionRAS-08 · DESC

Reprocessing partitions the system into three sterility classes. Mechanical hardware (halo, frame, distal shroud, needle holder) is steam-autoclaved at 134 °C for 18 min per ISO 17665. The sensor head and electronics shroud are VHP-compatible (STERRAD 100NX standard cycle) since steam is incompatible with the lens cement and CMOS package.

Single-use cartridges are gamma-irradiated at 25–40 kGy per ISO 11137 with verified SAL of 10⁻⁶. Each lot is dosimetry-mapped and the dose-mapping report is linked to the cartridge RFID payload so the robot rejects out-of-spec lots.

The sterile field is established with a draped, sealed end-effector shroud that mates to the kinematic coupling through a sterile silicone bellows; bellows leakage is monitored by a 0.5 kPa differential-pressure sensor.

Technical ParametersRAS-08 · TABLE
ParameterValueUnitTolerance / Note
Autoclave cycle134 °C / 18 minISO 17665, halo + frame + shroud
VHP cycleSTERRAD 100NX stdSensor head + electronics shroud
Gamma dose25–40kGyCartridges only
Validated SAL10⁻⁶Per ISO 11137 / 17665
Bellows leak threshold0.5kPa
Reprocessing cycles≥ 200Mechanical hardware lifecycle